最新中文字幕av无码专区不卡_67194成是人免费无码_久久视频热_亚洲黄色小说视频_亚洲www视频_影视先锋AV资源噜噜

RF Probe
Model 120 (75 to 110 GHz )
Model 120 (75 to 110 GHz / WR-10 waveguide input) 
  • Parameter
  • Download


High Performance Microwave Probes

Model 120

Features

  • Durable 75 GHz to 110 GHz
  • Insertion loss 1.75 db max (75 to 110 GHz)
  • Return loss 15 db max (75 to 110 GHz)
  • Individually spring loaded contacts
  • Measurement repeatability -50 db
  • Bias T option available
  • Patented coaxial design




undefined
FIGURE 1

The MODEL 120 PICOPROBE? sets new standards in microwave probing performance. Benefiting from coaxial techniques, which have inherent low loss and low dispersion characteristics, the Model 120 Picoprobe?, achieves an insertion loss of less than 1.75 db (typical) and a return loss of greater than 15 db (max.) over its frequency range.

With its individually spring loaded, Beryllium-Copper tips, the Model 120 Picoprobe? provides reliable contacts, even when probing non-planar structures. This reliable low resistance contact is one of the keys to providing highly repeatable measurements (-50 db) at W band frequencies. The Model 120 Picoprobe also provides direct viewing of the probe tips for accurate positioning.

Any pitch (tip spacing) from 50 microns to 450 may be specified. The probe can be configured with Ground-Signal-Ground (G,S,G), Ground-Signal (G,S), or Signal-Ground (S,G) tip footprints. We recommend smaller pitches with a G,S,G footprint for best performance.




120btl
left: Typical uncalibrated performance of a Model 120-GSG-150-BT while touching a 50 Ohm load on our CS-5 Calibration Substrate.
120btr
left: Typical insertion loss of a Model 120-GSG-150-BT.
120bloss
left: The Picoprobe? Model 120's Bias T provides a direct low resistance DC path for supplying up to 1.5 A to a device under test. The Bias T has special circuits which add loss at frequencies below the cut-off frequency (59.05 GHz) of WR-10 waveguide. This data was taken by launching a 40 MHz to 67 GHz signal from the tips of a calibrated Model 67A into the tips of a Model 120-GSG-150-BT. Without this low frequency loss, most active devices will oscillate.
120smith

left: Forward reflection of a Model 120 into an open ended 50 ps coplanar line after an LRM/TRM calibration was performed using our

CS-5 Calibration Substrate. The smooth inwardly spiraling line shows the increasing loss of the 50 ps coplanar line with frequency, coupled with a smoothly changing phase. The CS-5 Calibration Substrate will also perform excellent LRL/TRL and SOLT W-band calibrations. More data is available on request.

FIGURE 2

Micropositioners are available to hold Wiltron V band modules so that the module and the probe are micropositioned as a unit. In this way, total ease of positioning can be achieved with minimum insertion loss.
The probes also have a 3 hole mounting adaptor which will fit standard microwave probe stations. In this case, a longer section of waveguide may be desired to add positioning flexibility.
undefined
FIGURE 3
120lit6 (1)
Model 120-BT Waveguide Dimensions


Model 120-BT-M Waveguide Dimensions

FIGURE 4




Recommended Products
主站蜘蛛池模板: 国产精品视频色 | 久久久久人妻一区精品色欧美 | 亚洲午夜视频在线观看 | 国产午夜av片 | 91久久精品日日躁夜夜躁欧美 | 任你躁国产老女人 | 国产午夜av秒播在线观看 | 91精品国产99久久久久久久 | 爱草在线视频 | АⅤ资源天堂资源库在线 | 国产极品久久久久久久久 | 蜜桃视频在线观看一区 | 四虎永久免费 | 还珠格格第二部1至48集免费观看 | 亚洲国产日韩在线视频 | 一区二区三区国产 | 玩弄放荡人妻少妇系列视频 | A级毛片免费视频 | 国产日本三级 | 91精产品一区一区三区40p | 欧美日韩精品人妻狠狠躁免费视频 | 自拍论坛 | 精品久久久久久久久久久久久久久久 | 精品一区91 | 亚洲男人AV天堂午夜在 | 欧美精品无码久久久 | 在线观看视频中文字幕 | 狠狠爱亚洲综合久久 | 1—42集免费观看 | 亚洲色大成网站WWW在线观看 | 欧洲熟妇色XXXX欧美老妇免费 | 国产一级视频在线观看 | 麻豆网神马久久人鬼片 | 青青草草视频 | 人禽猛交免费观看 | 欧美VIDEO变态另类 | 91高清视频免费观看 | 边摸边脱边吃奶边做视频 | 国产99视频精品免费视看6 | 久久久久久一级毛片免费无遮挡 | 江苏少妇性BBB搡BBB爽爽爽 |